User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Power Designs 2005A:Sub 1 DATE: 12-Mar-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 102 CONFIGURATION: Datron 1281 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- P F 1.003 DISP Set the UUT front panel controls as follows: 1.003 DISP 100æV/DIV vernier .................................... 0 1.004 IEEE [@1281]*RST 1.005 CALL 2020B Sub 5 1.006 MATH M[1] = MEM 1.007 DISP Set the UUT front panel controls as follows: 1.007 DISP 100æ/DIV vernier ................................. .1mV 1.008 CALL 2020B Sub 5 1.009 MATH MEM = MEM - M[1] 1.010 ACC 200 .100mV 7P% 0.5P/ 1.011 MEME 1.012 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1001, Sys Tol 0.0001007, TUR 994.042 (>= 4.00). 2.001 DISP Set the UUT front panel controls as follows: 2.001 DISP 100æ/DIV vernier ................................. .2mV 2.002 CALL 2020B Sub 5 2.003 MATH MEM = MEM - M[1] 2.004 ACC 200 .200mV 7P% 0.5P/ 2.005 MEME 2.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1002, Sys Tol 0.0001014, TUR 988.166 (>= 4.00). 3.001 DISP Set the UUT front panel controls as follows: 3.001 DISP 100æ/DIV vernier ................................. .3mV 3.002 CALL 2020B Sub 5 3.003 MATH MEM = MEM - M[1] 3.004 ACC 200 .300mV 7P% 0.5P/ 3.005 MEME 3.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1003, Sys Tol 0.0001021, TUR 982.370 (>= 4.00). 4.001 DISP Set the UUT front panel controls as follows: 4.001 DISP 100æ/DIV vernier ................................. .4mV 4.002 CALL 2020B Sub 5 4.003 MATH MEM = MEM - M[1] 4.004 ACC 200 .400mV 7P% 0.5P/ 4.005 MEME 4.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1004, Sys Tol 0.0001028, TUR 976.654 (>= 4.00). 5.001 DISP Set the UUT front panel controls as follows: 5.001 DISP 100æ/DIV vernier ................................. .5mV 5.002 CALL 2020B Sub 5 5.003 MATH MEM = MEM - M[1] 5.004 ACC 200 .500mV 7P% 0.5P/ 5.005 MEME 5.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1005, Sys Tol 0.0001035, TUR 971.014 (>= 4.00). 6.001 DISP Set the UUT front panel controls as follows: 6.001 DISP 100æ/DIV vernier ................................. .6mV 6.002 CALL 2020B Sub 5 6.003 MATH MEM = MEM - M[1] 6.004 ACC 200 .600mV 7P% 0.5P/ 6.005 MEME 6.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1006, Sys Tol 0.0001042, TUR 965.451 (>= 4.00). 7.001 DISP Set the UUT front panel controls as follows: 7.001 DISP 100æ/DIV vernier ................................. .7mV 7.002 CALL 2020B Sub 5 7.003 MATH MEM = MEM - M[1] 7.004 ACC 200 .700mV 7P% 0.5P/ 7.005 MEME 7.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1007, Sys Tol 0.0001049, TUR 959.962 (>= 4.00). 8.001 DISP Set the UUT front panel controls as follows: 8.001 DISP 100æ/DIV vernier ................................. .8mV 8.002 CALL 2020B Sub 5 8.003 MATH MEM = MEM - M[1] 8.004 ACC 200 .800mV 7P% 0.5P/ 8.005 MEME 8.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1008, Sys Tol 0.0001056, TUR 954.545 (>= 4.00). 9.001 DISP Set the UUT front panel controls as follows: 9.001 DISP 100æ/DIV vernier ................................. .9mV 9.002 CALL 2020B Sub 5 9.003 MATH MEM = MEM - M[1] 9.004 ACC 200 .900mV 7P% 0.5P/ 9.005 MEME 9.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.1009, Sys Tol 0.0001063, TUR 949.200 (>= 4.00). 10.001 DISP Set the UUT front panel controls as follows: 10.001 DISP 100æ/DIV vernier .................................. 1mV 10.002 CALL 2020B Sub 5 10.003 MATH MEM = MEM - M[1] 10.004 ACC 200 1.000mV 7P% 0.5P/ 10.005 MEME 10.006 MEMC mV 0.1% 0.1U 1mVÿRange #! Test Tol 0.101, Sys Tol 0.000107, TUR 943.925 (>= 4.00). 11.001 DISP Set the UUT front panel controls as follows: 11.001 DISP 100æV/DIV vernier .................................... 0 11.002 END